• LPC-S101 Laser Dust Particle Counter
  • LPC-S101 Laser Dust Particle Counter

LPC-S101 Laser Dust Particle Counter

No.LPC-S101 2.83L
The LPC-S101 Type Laser Dust Particle Counter

Purpose: It is used to measure the size and quantity of dust particles per unit volume of air in a clean environment, and can also be used for testing accuracy testing.

Application Fields: Widely applied in departments such as integrated circuits
Particle size range::
0.1μm
0.15μm
0.2μm
0.25μm
0.3μm
0.5μm
1.0μm
5.0μm
  • LPC-S101 Laser Dust Particle Counter

Description

MODEL LPC-S101
Particle size range: Eight gears of 0.1μm、0.15μm、0.2μm、0.25μm、0.3μm、0.5μm、 1.0μm、5.0μm
Flow rate: 2.83L/min
Maximum concentration limit: ≥ 35000 particles/L @ 10% counting loss
Measurement cycle: Adjustable from 1s to 1000s
Minimum particle size range: 50 ± 20% @ 0.1 μm
Self-cleaning time: ≤ 10min
Storage environment: Temperature -15°C to 60°C, humidity 0 - 98% (no condensation)
Working environment: Temperature 5°C to 35°C, humidity 20 - 95% (no condensation)
Communication mode: RS232/RS485
Power supply: Lithium battery life is about 5 hours (adapter connected to 220V AC)
Data storage: At least 100,000 sets of data records
Shell: 304 stainless steel
Size: 200mm (length) × 180mm (width) × 160mm (depth) (excluding air nozzles, handles and other accessories)
Indication error: Not more than ± 20% FS